Cleaning and service
37
6.3.1.2 Remove the feed hopper
2.
Risk of injury to skin and hands
Fast rotating blades
– There is a risk of injury to hands, fingers and skin.
• Never operate the device without the feed hopper.
Fig. 25: Mounting the feed hopper
• Disconnect the device from the mains by pulling the plug from the socket.
• Open the grinding area door (T).
• Unscrew the three screws (TS).
• Lift the feed hopper out of the device for cleaning.
Follow the instructions in the following section to mount the feed hopper.
6.3.1.3 Mount the feed hopper
• Open the grinding area door (T).
• Place the feed hopper (R) on the device (see diagram).
It is possible for the feed hopper (R) to fall from the device as long
as it is not secured by the two socket screws.
• Screw in the two socket screws supplied(S).
• Initially only tighten the screws loosely.
• Close the grinding area door (T).
• Twist the rotary grip of the grinding area door.
• Align the feed hopper so that the plunger can be easily moved up and down
without tilting.
• Screw the three socket screws tight (10Nm).
Comments to this Manuals