Page view 28
Operating the machine
29
5.6 Inserting the filter unit and collecting receptacle
Pos: 7.22 /0 001 0 B edien ung sa nleit ung en K a pitels a mmlu nge n/SM 10 0/00 15 SM 1 00 B edi enu ng/1 565 SM 100 M od ul Ri ng sie b Mo ntag e @ 3\mod_1284973929211_9.doc @ 23764 @ @ 1
The filter unit serves as air outlet for the flow of air generated by the grinding
rotors.
Fig. 2: Attaching and removing the filter unit
Insert the bayonet lock (BV) on the filter unit (J) into the discharge flange
(AV) as shown in the diagram.
Twist the filter unit in a clockwise direction to engage the bayonet lock.
To remove the filter unit, pull the locking pins (RB) out to release the bayonet
lock (BV).
Alternatively the collecting receiver can be placed directly on the discharge flange.
The discharge of the air flow is prevented in the latching position (BV).
In the latching position (LS) a gap remains between the discharge flange and
collecting vessel; this permits air discharge.
Fig. 3: Collecting receiver and filter unit
Pos: 7.23 /0 000 5 Üb ers chrif te n/1. 1 Ü bersc hri ft en/ 1.1 Ü ber schr ift en BD A/ 11 E in- / Aussc halt en @ 0\mod_1229416527496_9.doc @ 5058 @ @ 1
5.7 Switching On and Off
Pos: 7. 24 /00 010 Bedi enung sanl eitung en Ka pitels ammlu ngen/ SM10 0/0015 SM100 B edien ung/1 510 SM 100 Mod ul Ein und Aussc halt en @ 3\mod_1284973915225_9.doc @ 23665 @ @ 1
BV
RB
AV
LS