Retsch AS 200 jet User Manual Page 44

  • Download
  • Add to my manuals
  • Print
  • Page
    / 56
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
Cleaning and service
44
Pos: 14.1 /0020 Üb erschriften/1 . Überschriften/1 Reinigung und Wartung @ 0\ mod_12311 67007723_9. docx @ 5450 @ 1 @ 1
7 Cleaning and service
Pos: 14.2 /0025 Warnhinweise/W 0003 WARNUN G Reinigung Stromstoß neu @ 1\mod_12 36239978437 _9.docx @ 7686 @ @ 1
WARNING
Risk of a fatal electric shock
- An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
Do not clean the blender under running water. Use only a cloth dampened
with water.
• Disconnect the power supply plug before cleaning the blender.
Pos: 14.3 /0001 B edienungsanlei tungen/001 06 Kapitelsam mlung AS200Jet/ 0010 AS200Jet Reinigung und Wartung/AS20 0Jet Modul R einigung und W artung @ 1\ mod_123563868 8043_9.docx @ 7479 @ 33 @ 1
7.1.1 Cleaning
We recommend Retsch ultrasonic baths for thorough, gentle and time-saving
cleaning of your test sieves.
Ask for our free special publication “Looking after and cleaning test sieves.
Fig. 22: Cleaning the air outlet channel
Clean the air outlet channel regularly using a brush to remove any deposits.
7.1.2 Vacuum cleaner changing the vacuum cleaner bag
Change the vacuum cleaner bag or empty the dust container on your vacuum
cleaner regularly as required.
Check the degree of soiling of the vacuum cleaner filter regularly and change the
filter where applicable.
The vacuum or differential pressure generated by the vacuum cleaner may
otherwise be too low for sieving.
Pos: 14.4 /0001 B edienungsanlei tungen/00106 K apitelsammlung AS200Jet/0 010 AS200Jet Rei nigung und Wartung/AS20 0Jet Modul Wartu ng @ 1\mo d_124999235 2900_9.doc x @ 13778 @ 2 @ 1
7.1.3 Maintenance
This device is essentially maintenance-free if cleaned regularly.
It is possible to calibrate the differential pressure sensor within the context of the
DIN EN ISO 9001:2000 ff. Please contact your dealer or Retsch GmbH.
www.retsch.com
Pos: 14.5 /0020 Üb erschriften/1 .1 Überschriften/ 1.1 Überschrif ten BDA/11 Kali brieren des Druc ksensors (überpr üfen des Diff erenzdruc kes ) @ 2\mod_ 12779715140 50_9.docx @ 205 28 @ 2 @ 1
7.2 Calibrating the pressure sensor (checking of the differential pressure)
Pos: 14.6 /0005 RET SCH/0005 R ETSCH Be dienungsanleitunge n Kapitelsamml ungen/AS20 0Jet/0010 AS200J et Reinigung und Wartung/AS 200Jet Mo dul Kalibrieru ng @ 2\mod_127 7967985047_ 9.docx @ 2052 0 @ @ 1
Unscrew the sealing screws (MS).