15.11.2006 Retsch GmbH 11 Doc. Nr. D 98.725.9998
Temperature variations
If temperature variations are high (e.g. during air
transport) the RS200 must be protected against
condensed water. Otherwise the electronic components
can become damaged.
Intermediate storage
Ensure that the RS200 is also stored dry during
intermediate storage.
Erection
Erect the RS200 on a firm base. Further parameters are given
in the section „Technical data”.
Net weight of RS200: approx. 210 kg
The machine must be locked in position before starting up.
For this purpose press the locking lever F of the two front
rollers downwards.
Parameters for the place of installation
Ambient temperature: 5°C to 40°C
If the ambient temperature drops below or exceeds these
values the electrical and mechanical components can become
damaged and performance data are changed to an unknown
extent.
Atmospheric humidity
Maximum relative humidity = 80% at temperatures up to 31°C,
decreasing linearly down to 50% relative humidity at 40°C.
At higher atmospheric humidity the electrical and mechanical
components can become damaged, and performance data are
changed to an unknown extent.
Installation height: max. 2000 m above sea level
Transport safety device
The driving unit of the RS200 is locked for transport.
Before starting up for the first time the two cheese-head
screws Z fitted under the rubber cover G must be
removed (Fig. 3 Transport safety device).
To do this lift the rubber cover G upwards and unscrew
the cheese-head screws. Then pull out the cheese-head
screws side wards together with the sleeve P (Fig. 4
Removing the transport safety device).
Keep the cheese-head screws for further transport.
Z
Z
G
Fig. 3 Transport safety device