Retsch TG 200 User Manual Page 5

  • Download
  • Add to my manuals
  • Print
  • Page
    / 28
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 4
2/25/2013 Retsch GmbH 5 0002
Safety instructions, summary, part 1
Safety instructions
We exclude any liability for claims for damages in any form whatsoever for injury to persons or damage to the
machine caused through non-observance of the following safety instructions.
Use according to the intended purpose
Do not make any alterations to the machine and use only spare parts and accessories approved by Retsch.
Otherwise the Declaration of Conformity with European directives made by Retsch will lose its validity.
Furthermore this will also lead to the loss of any guarantee claims.
Do not dry explosives or easily flammable materials, or materials with a flashpoint below 180°C in the TG200.
Risk of fire and explosion.
Do not operate the TG200 in the vicinity of easily flammable materials.
Risk of fire and explosion.
You must check to what extent toxic danger exists through the charged material.
Risk to health through vapours or residual dusts which are not retained by the filters.
Do not operate the TG200 without supervision.
Risk of fire and explosion.
Materials should be dried only in the appropriate time, at the appropriate temperature. Take into consideration the
melting point and temperature sensitivity of the material.
The material to be dried can be destroyed through overheating.
Packing materials
Please keep the packing material for the duration of the guarantee period, since, in case of a complaint and return
of the machine in unsuitable packing material your guarantee claim will be at risk.
Transport
The TG200 must not be jolted, shaken or thrown during transport. Otherwise the electronic and mechanical
components may be damaged.
Temperature variations
If subjected to high temperature variations (e.g. during transport by aircraft) the TG200 must be protected
against condensed water. Otherwise the electronic components may be damaged.
Supply schedule
If the delivery is incomplete and/or transport damaged is found, you must inform the transporter and Retsch
GmbH immediately (within 24 hrs). Later complaints may possibly be no longer considered.
Ambient temperature:
If the temperature drops below, or exceeds the ambient temperature the electrical and mechanical components
may be damaged; the performance data are then altered to an unknown extent.
Atmospheric humidity:
At high atmospheric humidity the electrical and mechanical components may be damaged and the performance
data are altered to an unknown extent.
Installation
Maintain a safety distance of 100 mm all round.
The hot air outlet can ignite easily flammable objects.
Risk of fire and explosion.
Electrical connection / connecting the power supply
If the values on the rating plate are not observed the electrical and mechanical components may be damaged.
Page view 4