Retsch TG 200 User Manual Page 11

  • Download
  • Add to my manuals
  • Print
  • Page
    / 28
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 10
2/25/2013 Retsch GmbH 11 0002
Parameters for the place of installation
Ambient temperature:
5°C to 40°C
If the temperature drops below or exceeds the ambient
temperature the electrical and mechanical components may
be damaged. The performance data are altered to an
unknown extent.
Atmospheric humidity:
Maximum relative humidity 80% at temperatures up to 31°C, linear
decrease up to 50% relative humidity at 40°C.
At high atmospheric humidity the electrical and mechanical
components may be damaged. The performance data are
altered to an unknown extent.
Installation height:
Max. 2000 m above sea level.
Installation
Install the TG200 only on a robust laboratory bench.
Maintain a safety distance of 100 mm all round.
The hot air outlet can ignite easily flammable objects.
Risk of fire and explosion.
Electrical connection
The voltage and frequency of the TG200 is given on the
rating plate.
Ensure that the values agree with the existing mains
supply.
Connect the TG200 to the mains supply using the
connection cable supplied with the machine.
An external fuse of min. 20A should be installed when
connecting the power supply cable to the mains, in
accordance with the regulations at the place of
installation.
If the values on the rating plate are not observed the
electrical and mechanical components may be damaged.