08.02.2012 Retsch GmbH 28 0003
F13 to F26
The display shows
The display shows
F13
F14
MOTOR IS COMING
TO A STAND STILL
FAULT SPEED SENSOR
SERVICE REQUIRED!
F16
The display shows
The display shows
F15
PROBLEM WITH
MOTOR HAS OVERHEATED
THE FREQUENCY
NO START POSSIBLE
CONVERTER
LET MOTOR COOL DOWN
SERVICE REQUIRED!
The error appears when the unit
has overheated and yet START is
pressed.
The display shows
The display shows
F17
F18
MOTOR HAS OVERHEATED
PROBLEM
WAS SWITCHED OFF
WITH THE
LET MOTOR COOL DOWN
TRANSFORMER
CONTINUE WITH STOP
SERVICE REQUIRED!
The error appears when the
critical motor temperature is
reached during grinding.
F26
The display shows
FREQUENCY CONVERTER
HAS OVERHEATED
NO START POSSIBLE
ALLOW TO COOL DOWN
The error appears when the
frequency converter has
overheated and yet START is
pressed.