31.07.2012 Retsch GmbH 8
SM
Set up the AS200 only on a stable laboratory bench to avoid un-
pleasant transmission of vibrations.
*
Loosen the hexagonal nuts SM using an open-ended
spanner until the sieve plate moves freely
Unscrew the wing screws A (shipping nuts) with hexagonal
nuts.
Mount the appropriate sieve clamping device (See the
chapter on operation for instructions)
Keep the wing screws A with hexagonal nuts (shipping nuts)
in a safe place for later transportation.
Reusing the shipping nuts
Screw the hexagonal nuts fully onto the wing screws
Screw the wing screws through the sieve plate into the
housing
Twist the hexagonal nuts downwards and tighten until the
sieve plate can no longer move
Operation with these shipping nuts in place, or shipping without
the these nuts installed, could cause damage to mechanical
components.
Electrical connection
* Refer to the data plate for information on the voltage and
frequency at which the AS200 is to be operated.
* Ensure that the values shown there correspond to those for
the local power supply.
* Use the supplied power cord to connect the AS200 to the
power supply.
* When connecting the power cord to the power source, be sure
that the supply circuit is fused in accordance with local codes.
Failure to observe the values on the data plate can cause
damage to either the electrical or the mechanical com-
ponents or both.